The custom RFPD had non-stationary noise that would briefly spike up across about a 1 MHz range by 20-30dB above the usual PD noise level. This was fixed by modifying the RFPD's power cable so that the power voltages and PD bias voltage were picked off of the same cable about 1cm before the PD, instead of about 1 meter before the PD at a patch panel. I believe this was creating a loop between the ground cables and possibly also between the different -15V cables. (The power supply cable has +15V, GND, -15V and the bias cable has -15V and GND.) The other custom PDs suffer from excessive 60 Hz noise, and re-soldering cables for them will likely cut this noise. This doesn't seem like an essential thing to do in the short term, but should be fixed in a later version of the squeezer and possibly for the PDs used to lock the SHG and OPO cavities.