ADBC table floating from N2 tank
Myron plumbed the ADBC table to a nitrogen tank rather than the building air supply. This is a size 300 cylinder.

Valve open at 10 am. The tank pressure started around 2500 psi. Now at 5 pm it is still a little above 2400 psi. The table work today was minimal.

The regulator on the tank reads 45 psi. The gauges on the table legs read more like 15–25 psi.

These are S-2000 series Newport legs.
Comments related to this report
evanhall - 17:26, Thursday 25 May 2023 (11206)
At 5 pm May 25, the tank pressure was just over 2300 psi. The regulator pressure was 44 psi and the table legs were again at 15–20 psi.

The table floats but I suspect not as much as it ultimately should — it seems a bit more sluggish than when we were drawing from the building air at 60 psi. The table is Newport RS-4000-48-12, weight 1110 pounds. The two vacuum tanks, pipe, and inner optical platforms all told are probably north of 700 lbs. The S-2000 manual gives the recommended leg pressure as (Load / (4 × 25.9 sq. in.)) + 10 psi, which for us comes out to 22 psi at least.

It means we may want to repeat this cylinder test with a higher leg pressure. That may require a tank regulator with higher range.
evanhall - 11:26, Tuesday 30 May 2023 (11212)
At 6 am May 30, the tank pressure is about 1990 psi and the regulator pressure is 44 psi. The average tank depletion rate is therefore about 110 psi per day.
evanhall - 17:02, Sunday 18 June 2023 (11229)

On June 18, the regulator pressure is 44 psi and the tank pressure is about 1020 psi.