Displaying report 1-1 of 1.
Reports until 18:19, Tuesday 11 April 2023
GRAVITES General
epolini - posted 18:19, Tuesday 11 April 2023 (11197)
Rio fiber laser operational in NW22
Today, with Xinghui and Dorotea, we connected the RIO fiber laser on the bench on NW22. We connected the temperature and current stabilization drivers to the laser, using the movable rack in fig. 1. Everything is working properly. We need to be careful to set an output power lower than 10mW. We engaged the temperature stabilization with the settings shown in fig. 2.
NB. the internal interlock light on the laser is working only if the temperature stabilization is activated BUT IT IS EMITTING EVEN IF THE LIGHT IS OFF (see fig. 3). 

Since this laser is not connected to the lab interlock, WE MUST TURN OFF THE LASER WHEN WE LEAVE THE LAB.

Images attached to this report