Today I replaced the 3 fibers with a new one ordered on thorlabs (SM980-5.8-125-CUSTOM) 15 meters long. I measured power losses before and after. The power on the powermeter was fluctuating, so I considered the highest detected power. BEFORE the replacement: Power on the bench before the first fiber (after the polarization tuning) : 38 mW Power after the 1st fiber : around 38 mW Power at the entrance of the ifo : 28 mW AFTER the replacement I decided to connect the fiber before the polarization tuning Power on the bench before coupling in the new fiber : 50 mW Power at the end of the fiber, before the ifo : 45 mW I left two x 1 meter fibers next to PDs, in order to replace the PM fibers with SM fibers. On the bench in green lab there is also a 10 meters fiber without jacket to replace the arm of the ifo. Next steps: replace the fibers and place the ifo on two small stages at different heights to start thinking about the vertical switches and isolation stages.